 Agilent E3640A - E3649A Programmable DC Power Supplies
Data Sheet
Reliable Power, Repeatable Results Single and dual outputs Dual range output 30 W to 100 W output power Front and rear output terminals Over-voltage protection Remote Sensing GPIB and RS-232 standard Save and recall functions
Great Performance, Outstanding Price
With the output power of 30 to 100 W, the Agilent E364xA Series programmable DC power supplies provide great performance at a great price. All ten models deliver clean power, excellent regulation, fast transient response and built-in GPIB and RS-232 interfaces. They are designed to meet the needs of R&D design verification, production testing, QA verifications, and other demanding applications with Agilent Technologies's quality and reliability.
Remote Interface
Agilent E364xA power supplies support any PC with a GPIB (IEEE-488) card or RS-232 interface. Every model ships standard with both GPIB and RS-232. The easy-to-use SCPI (Standard Commands for Programmable Instruments) allow fast and simple programming procedures. Besides, the user manual provides sufficient information on programming for all end users, from beginners to veterans.
The output on/off button sets the output to zero. If you own a dual output model, you can view two voltages or currents that are displayed simultaneously.
Versatile Power
Agilent E364xA power supplies give you the flexibility to select from dual output ranges. Output load is protected against overvoltage, which can be easily monitored and adjusted from the front panel and remote interface. Remote sensing is available in the rear terminal to eliminate errors caused by voltage drops on the load leads. These power supplies offer new versatile binding posts on the front panel and screw-type terminals on the rear panel. New front panel binding posts allow you to use safety test leads as well as conventional banana clips and stripped wires. An optional rackmount kit is available. The Agilent E364xA Series employs a cooling fan with automatic speed control to reduce the acoustic noise. LabView and LabWindows are registered trademarks of National Instruments. Microsoft Windows 98 and Windows NT are US registered trademerks of Microsoft Corporation.
Broad Support
VXI plug and play software drivers are available for Agilent VEE, National Instruments LabViewTM and LabWindowsTM. With these drivers, integration of the E364xA into your system can never be any easier. The drivers are supported under Microsoft(R) Windows 98(R) and NT(R).
Steady Output
With 0.01 percent load and line regulation, Agilent E364xA power supplies are able to maintain a steady output when power line and load changes occur. They also specify normal mode voltage noise and low common mode current noise. The low normal mode noise specification assures clean power for precision circuitry applications, and the low common mode current provides isolation from power line current injection. Agilent E364xA power supplies specify less than 90 msec of voltage settling time at any output load condition.
Front Panel Operation
An easy-to-use rotary knob and selfguiding keypads allow you to set the output at your desired resolution without any effort. Also, both voltage and current levels can be set to a maximum resolution of 10 mV/1 mA from the front panel. In addition, you can store and recall for up to five complete power supply setups using the internal non-volatile memory.
2
Set the overvoltage protection and enable the overvoltage protection circuit Highly visible vacuum flourescent display Rotary knob for quick and analog-like control of voltage and current Tough handle for easy carriage and unit prop-up Versatile binding posts offer flexibility to use safety test leads, banana plugs or stripped wires
Selectable dual range provides flexibility and convenience Store and recall up to five different operating statuses in nonvolatile storage locations
Electronic calibration from the front panel
1 year warranty protects your investment
Cooling fan with automatic speed control for low acoustic noise Built-in GPIB and RS-232 interfaces Rear output and sensing terminals
3
Agilent E3640A - E3649A Programmable DC Power Supply Specifications
Model Number Maximum Power # of Outputs DC Output Rating (@ 0 C to 40 C) Net Weight Dimensions (without bumper) Model Number Maximum Power # of Outputs DC Output Rating (@ 0 C to 40 C) Net Weight Dimensions (without bumper) Load and Line Regulation (% of output + offset) Voltage Current Ripple and Noise (20 Hz to 20 MHz) Normal Mode Voltage Normal Mode Current Common Mode Current Accuracy1 12 Months (@ 25 C 5 C), (% output + offset) Programming Voltage Current Readback (over GPIB and with respect to actual output) Voltage Current Meter2 (over front panel with respect to actual output) Voltage Current Resolution Program Readback Meter Transient Response Settling Time3 OVP Accuracy, (% output + offset) Activation Time4
1 2
E3640A 30 W 1 0 to 8 V/3 A or 0 to 20 V/1.5 A 5.3 kg
E3641A
E3642A 50 W
E3643A
E3644A 80 W
E3645A
1 0 to 35 V/0.8 A or 0 to 60 V/0.5 A 5.2 kg
1 0 to 8 V/5 A or 0 to 20 V/2.5 A 6.3 kg
1 0 to 35 V/1.4 A or 0 to 60 V/0.8 A 6.2 kg
1 0 to 8 V/8 A or 0 to 20 V/4 A 6.6 kg
1 0 to 35 V/2.2 A or 0 to 60 V/1.3 A 6.7 kg
212.6 mm W x 88.5 mm H x 348.3 mm D E3646A 60 W 2 Two 0 to 8 V/3 A or 0 to 20 V/1.5 A 8.2 kg 2 Two 0 to 35 V/0.8 A or 0 to 60 V/0.5 A 8.0 kg 2 Two 0 to 8 V/5 A or 0 to 20 V/2.5 A 9.2 kg E3647A E3648A 100 W 2 Two 0 to 35 V/1.4 A or 0 to 60 V/0.8 A 9.1 kg E3649A
212.8 mm W x 133.0 mm H x 348.3 mm D
< 0.01% + 3 mV < 0.01% + 250 A < 5 mVpp/0.5 mVrms for 8 V/20 V models < 8 mVpp/1 mVrms for 35 V/60 V models < 4 mArms < 1.5 Arms
< 0.05% + 10 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A) < 0.2% + 10 mA < 0.05% + 5 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A) < 0.15% + 5 mA (< 0.15% + 10 mA for output 2 of E3646/47/48/49 A) < 0.05% + 2 counts (<0.1% + 4 counts for output 2) < 0.15% + 5 mA (<0.15% + 10 mA for output 2) < 5 mV/1 mA < 2 mV/1 mA 10 mV/1 mA Less than 50 sec for output to recover to within 15 mV following a change in output current from full load to half load or vice versa. < 90 msec < 0.5% + 0.5 V < 1.5 msec, OVP 3 V/< 10 msec, OVP < 3 V
Accuracy specifications are valid after a 1-hour warm-up with no load and calibration at 25 C. Meter accuracy specification is at minimum 10 mV decimal limited by front panel resolution. 3 Maximum time required for the output voltage to change from 1% to 99% or vice versa following the receipt of VOLTage or APPLy command via direct GPIB or RS-232 interface. 4 Average time for output to start and drop after an OVP condition occurred.
4
Temperature Coefficient per C (% output + offset) Voltage Current Stability, constant load & temperature (% output + offset), 8 hrs Voltage Current Remote Sensing Max. voltage drop in each load lead AC Input (47 Hz - 63 Hz) Warranty Product Regulation < 0.02% + 2 mV < 0.1% + 1 mA 1V 100 Vac 10% (Opt 0E9)/115 Vac 10% (Std)/230 Vac 10% (0E3) One year for E364xA series power supplies Three months for standard shipped accessories Designed to comply with UL3111-1; certified to CSA 22.2 No. 1010.1; conforms to IEC 1010-1; complies with EMC directive 89/336/EEC (Group 1, Class A) < 0.01% + 3 mV (< 0.02% + 5 mV for output 2 of E3646/47/48/49A) < 0.02% + 3 mA
Ordering Information
Agilent E364xA Series Power Supplies E3640A 30-Watt Single Power Supply E3641A 30-Watt Single Power Supply E3642A 50-Watt Single Power Supply E3643A 50-Watt Single Power Supply E3644A 80-Watt Single Power Supply E3645A 80-Watt Single Power Supply E3646A 60-Watt Dual Power Supply E3647A 60-Watt Dual Power Supply E3648A 100-Watt Dual Power Supply E3649A 100-Watt Dual Power Supply
Rackmount Kits*
Agilent E3640A/41A/42A/43A/44A/45A To rackmount two instruments side-by-side Lock-link Kit (P/N 5061-9694) Flange Kit (P/N 5063-9212) To rackmount one or two instruments in a sliding support shelf Support Shelf (P/N 5063-9255) Slide Kit (P/N 1494-0015) required for support shelf For a single instrument, also order Filter Panel (P/N 5002-3999) Agilent E3646A/47A/48A/49A To rackmount two instruments side-by-side Lock-link Kit (P/N 5061-9694) Flange Kit (P/N 5063-9214) To rackmount two instruments in a sliding support shelf Support Shelf (P/N 5063-9256) Slide Kit (P/N 1494-0015) * Rackmounting with 1CM or lock-link/ flange kit requires Agilent or customer support rails Agilent Support Rails-E3663AC
Standard Shipped Accessories
User's & Service guide, Product Reference CD, AC power cord
Power Options
Opt. 0E3 230 Vac 10% Opt. 0EM 115 Vac 10% Opt. 0E9 100 Vac 10%
Other Options
Opt. 1CM Rackmount kit* - Single Output (P/N 5063-9240) - Dual Output (P/N 5063-9243) Opt. 0L2 Extra Manual Opt. UK6 Commercial calibration with test result data E3600A-100 Test lead kit
5
www.agilent.com
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Remove all doubt
Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to
www.agilent.com/find/removealldoubt
For more information on Agilent Technologies' products, applications or services, please contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus Phone or Fax Americas Canada Latin America United States Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand Europe & Middle East Austria Belgium Denmark Finland France
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test equipment solutions with confidence.
(877) 894-4414 305 269 7500 (800) 829-4444
www.agilent.com/find/open
Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of systemready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development.
is the US registered trademark of the LXI Consortium.
1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008
Product specifications and descriptions in this document subject to change without notice.
01 36027 71571 32 (0) 2 404 93 40 45 70 13 15 15 358 (0) 10 855 2100 0825 010 700*
*0.125 /minute
07031 464 6333 Germany 1890 924 204 Ireland 972-3-9288-504/544 Israel 39 02 92 60 8484 Italy 31 (0) 20 547 2111 Netherlands 34 (91) 631 3300 Spain 0200-88 22 55 Sweden 0800 80 53 53 Switzerland 44 (0) 118 9276201 United Kingdom Other European Countries: www.agilent.com/find/contactus
Revised: October 6, 2008
(c) Agilent Technologies, Inc. 2009 Printed in USA, July 29, 2009 5968-7355EN


▲Up To Search▲   

 
Price & Availability of E3646A

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X